The 3001 Bench Calibrator
combines the power and features of a high accuracy lab
calibrator,
with the addition of an isolated measurement channel.
This isolated measurement channel consists of two (2) voltage
ranges: 10V and 100V DC,
current range 0 to 52 mA, current range with simultaneous 24VDC
power, selectable 250 Ohm HART resistor
and accuracy of 0.005% of reading on voltage ranges.
3001 Bench Laboratory Calibrator
Model | Description |
---|---|
3001 | Bench Calibrator with Ultra-High Accuracy include Instruction Manual, power cord, and NIST Calibration Certificate. Test leads are NOT included. |
Superior calibration accuracy to 0.0025% of reading
Direct keyboard entry or cursor entry with decade control
Automatic standby function protects device under test
Source current, voltage, T/C, RTD (Ohms)
Read T/C, RTD and pressure
Source/read 11 thermocouple types
Source/read nine standard RTD types
Custom RTD/SPRT profile
Nine set points for each output range and type
Local or RS232 remote control
IEEE-488 control interface.
Compatible with Fluke Met/Cal® software
Isolated measurement channel
- Two (2) voltage ranges: 10V and 100 V DC
- MilliAmp range 0 to 52 mA
- MilliAmp range with simultaneous 24 VDC power
- Selectable 250 Ohm HART™ resistor
- Accuracy of 0.005% of reading on voltage ranges
Power Requirements: | 90/240 VAC |
Remote Interface: | RS-232 and GPIB (IEEE488-2 optional) |
Circuit Protection: | Internal self resetting fuses |
Display: | 16 x 2 large character alphanumeric with LED backlight |
Size: | Approx. 11.5” x 4.7” x 12” |
Weight: | Approx. 5 pounds |
Function | Output Range | Input Range | Resolution | Accuracy |
100mV | 0 to 100.000mV | N/A | 1µV | ± 0.003%, ± 3 µV |
1V | 0 to 1.00000V | N/A | 10µV | ± 0.003%, ± 10 µV |
10V | 0 to 10.0000V | N/A | 100µV | ± 0.003%, ± 100 µV |
100V | 0 to 100.000V | N/A | 1mV | ± 0.003%, ± 1 mV |
100mA | 0 to 100.000mA | N/A | 1µA | ± 0.005%, ± 1count |
T/C | J,K,T,E,R, S, N, B, L, U, C, BP, XK, mV |
J,K,T,E,R, S, N, B, L, U, C, BP, XK, mV |
Out 0.1 °
C/°
F In 0.01 ° C/° F |
Type J Typical 0.14°
C Type J Typical 0.14° C |
RTD | PT385 (100, 200, 500, 1000) PT392, PT3916 (JIS) Ni120, Cu10, YSI400 |
PT385 (100, 200, 500, 1000) PT392, PT3916 (JIS) Ni120, Cu10, YSI400, 25ohm SPRT |
Typical PT385-100 | |
Out 0.1 °
C/°
F In 0.001 ° C/° F |
±
0.10 °
C ± 0.020 ° C |
|||
ohms | 5 to 4000.0 ohm | 0 to 4000.0 ohm | ||
5 to 400.00 ohm | 0 to 400.00 ohm | Out 0.001Ω In 0.001Ω |
±
0.05Ω 40 PPM ± 0.002Ω |
|
5 to 4000.0 ohm | 0 to 4000.0 ohm | Out 0.01Ω In 0.01Ω |
±
0.3Ω 40 PPM ± 0.02Ω |
|
Pressure | 0-1" H2O to 10,000 psi | Module Dependent |
0.025% F.S. Typical |
|
Isolated Measurement Channel | ||||
Range | Accuracy | |||
0 - 10.0000 V | ± 0.005%, ± 0.2 mV | |||
0 - 100.000 V | ± 0.005%, ± 2.0 mV | |||
0 - 52.000 mA | ± 0.01%, ± 1 µV | |||
Loop Power | 24 V ± 10% | |||
HART resistor | 250 ± 3% | |||
Maximum Current | 24 mA | |||
Notes: 1) All W/RTD inputs
are 4W. 2) All accuracy specs are "% of reading” unless otherwise noted. |