The LC100 Current Loop Calibrator
provides significantly extended performance when
compared to any competitive calibrator.
With an accuracy of 0.015% of reading, with
0.001 mA resolution, the LC-100 has the highest accuracy in its
class.
Five preset outputs (4, 8, 12, 16, 20 mA)
Fuseless input protection up to 250 VAC
LC100 Current Loop Calibrator
Model | Description |
---|---|
LC100 | Current Loop Calibrator w/ knob adjust function |
Very high accuracy - ±
0.015% of reading
Very high resolution - 0.001 mA
% Error function - eliminates manual error calculation
Extended adjustment range—digital knob and user-selectable
decade output allow both large
and incremental (0.001 mA step) output changes
Reads 0 to 28.000 VDC
Built-in 250 Ohm resistor facilitates calibration of Hart™
devices
Remote calibration through Automatic Step and Ramp functions
Loop power measurement capability
Simulates, powers, and measures two-wire transmitters
Unique features, such as a “% Error”
function, which eliminates manual error calculation and
allows
the
display of the actual versus ideal error at any calibration
point, put the LC-100 way
beyond similar instruments.
The LC-100 calibrator can simulate, power, and measure two-wire
transmitters. With automatic
Step and
Ramp functions, the LC-100 loop calibrator enables remote
calibration of 4-20 mA devices.
Input | |
Current Range | 0.000 to 24.000 mA -25.00 to + 125.00 % |
Voltage reading | 0 to 28 VDC |
Input protection | Fuseless, up to 250 VAC |
Output | |
Source/Simulate Range | 0.000 to 24.000 mA -25.00 to +125.00 % |
Drive Capability | 1200 Ohm without HART 950 Ohm with HART |
Loop Supply Voltage | 24 VDC |
Range Select | Decade Incremental to 0.001 mA steps |
Resolution | 1 µA |
Accuracy | ± 0.015% of reading ± 2µA |
Temperature Range | |
Operating | -10 to + 55 degC |
Storage | -20 to + 70 degC |
Power Requirements | |
Battery | 9V one required |
AC Adapter | included |
Mechanical | |
Dimensions | 5.7" x 3.15" x 1.43" |
Weight | 12 ounces (340 grams) |