HD2106.1 Conductivity Meter Thermometer complete with combined conductivity temperature probe SP06T,
conductivity solution 12.880μS. Storage of maximum, minimum, average value, RS232 C output for data
transfer in real time to a PC or printer. Functions: REL, HOLD and auto power off which can be disabled,
IP 67 protection degree.
HD2106.2 Conductivity Meter Thermometer complete with combined conductivity/temperature probe SP06T,
conductivity solution 12.880μS. Datalogger which can store the maximum, minimum, average value and can
store up to 36.000 samples.RS232 C/ USB output for data transfer in real time to a PC or printer.
Functions: REL, HOLD and auto power off which can be disabled, IP 67 protection degree.
HD2106.1 Conductivity Meter
Model | Description |
---|---|
HD2106.1 | HD2106.1 Conductivity Meter Thermometer |
HD2106.2 | HD2106.2 Conductivity Meter Thermometer Data logger |
The HD2106.1 and HD2106.2 are portable instruments with a large LCD display.
They measure conductivity, liquid resistivity, total dissolved solids (TDS), and salinity using
combined
4-ring and 2-ring conductivity/temperature probes.
Temperature only is measured by Pt100 or Pt1000
immersion, penetration or contact probes.
The HD2106.2 is a datalogger. It memorizes up to 36,000 conductivity and temperature samples which can
be transferred from the instrument connected to a PC via the multi-standard RS232C serial port and USB 2.0.
The storing interval, printing, and baud rate can be configured using the menu.
The HD2106.1 and HD2106.2 models are fitted with an RS232C serial port and can transfer the acquired
measurements to a PC or to a portable printer in real time.
Measurement of conductivity by Instrument | |
Measurement range | (K cell=0.1) |
Resolution | 0.00 - 19.99μS7cm / 0.01μS/cm
(with K cell=0.1)
0.0 - 199.9μS/cm / 0.1μS/cm 200 - 1999μS/cm / 1μS/cm 2.00 - 19.99mS/cm / 0.01mS/cm 20.0 - 199.9mS /cm / 0.1mS/cm |
Accuracy (conductivity) | ±0.5% ±1digit |
Measurement of resistivity by Instrument | |
Measurement range / Resolution | 4.0 - 199.9Ω / 0.1Ω
200 - 999Ω / 1Ω 1.00k - 19.99kΩ / 0.01kΩ 20.0k - 99.9kΩ / 0.1kΩ 100k - 999kΩ / 1kΩ 1 - 10MΩ / 1MΩ |
Accuracy (resistivity) | ±0.5% ±1digit |
Measurement of total dissolved solids (with coefficient χ/TDS=0.5) | |
Measurement range (K cell=1) | 0.00 - 19.99mg/l / 0.05mg/l |
Resolution | 0.0 - 199.9 mg/l / 0.5 mg/l (with K cell=0.1)
200 - 1999 mg/l / 1 mg/l 2.00 - 19.99 g/l / 0.01 g/l 20.0 - 199.9 g/l / 0.1 g/l |
Accuracy (total dissolved solids) | ±0.5% ±1digit |
Measurement of salinity | |
Measurement range / Resolution | 0.000 - 1.999g/l / 1mg/l
2.00 - 19.99g/l / 10mg/l |
Accuracy (total dissolved solids) | ±0.5% ±1digit |
Measurement of temperature by Instrument | |
Pt100 measurement range | -50 - +200°C |
Pt1000 measurement range | -50 - +200°C |
Resolution | 0.1°C |
Accuracy | ±0.25°C |
Drift after 1 year | 0.1°C/year |
HD 2106.1 Conductivity-Thermometer measures conductivity, liquid resistivity, total dissolved solids, salinity
and temperature.
Storage of maximum, minimum, average value, RS232 C output for data transfer in real time
to a PC or printer.
Functions: REL, HOLD and auto power off which can be disabled, IP 67 protection degree.
Supplied with carrying
case, 4 batteries, instruction manual, Software DeltaLog 9. Conductivity probes,
calibrating solutions, TP47…
series temperature probes with SICRAM module and cables for data download
have to be ordered separately.
HD 2106.2 Conductivity-Thermometer measures conductivity, liquid resistivity, total dissolved solids, salinity
and temperature.
Datalogger which can store the maximum, minimum, average value and can store up to
36,000 samples. RS232 C/
USB output for data transfer in real time to a PC or printer. Functions: REL, HOLD
and auto power off which can be
disabled, IP 67 protection degree. Supplied with carrying case, instruction
manual, Software DeltaLog 9. Conductivity
probes, calibrating solutions, TP47… series temperature probes
with SICRAM module and cables for data
download have to be ordered separately.