HD32.8.16 Data Logger Temperature with display. 16 inputs thermocouple types K, J, T, N, R, S, B, E.
The HD32.8.8 and HD32.8.16 instruments are rugged data loggers.
The former has 8 inputs, the latter has 16 inputs.
They work with thermocouple probes type K, J, T, N, R, S, B and E with miniature connector.
HD32.8.8 Data Logger
Model | Description |
---|---|
HD32.8.8 | Data Logger Temperature with display. 8 inputs |
HD32.8.16 | Data Logger Temperature with display. 16 inputs |
Flash memory, organized in 64 blocks, for a total capacity of 800,000 recordings divided among the
existing inputs.
The recording can be handled in two ways:
- when the available memory is full, the collected data are overwritten
starting from the oldest
ones (circular memory),
- the recording stops when the available memory is full.
DeltaLog9 software, for data download and processing.
LCD backlit graphic display 128x64 pixel.
Instrument setup through the keyboard; no connection required to the PC.
Security password for keyboard locking.
Power supply: 4 1.5V alkaline C-BABY type batteries, external power supply
12VDC-1A or PC USB port.
Consumption @6VDC:
<60μA when the instrument is off
<60μA in sleep mode with all probes connected
<40mA during data logging with all probes connected
Use of the HD32.8.8 and HD32.8.16 data loggers: in the field for complex machine or equipment multi measurements, machine testing, pharmaceutical and food industry, oven mapping,
air conditioning units, etc.
Number of inputs | 8 for the HD32.8.8 16 for the HD32.8.16 |
Connection | Miniature female socket for thermocouple |
Instrument accuracy and measurement range |
Tc: K -200 - +1,370°C / ±0.1°C up to
600°C ±0.2°C over 600°C Tc: J -100 - +750°C / ±0.1°C up to 400°C ±0.2°C over 400°C Tc: T -200 - +400°C / ±0.1°C Tc: N -200 - +1,300°C / ±0.1°C up to 600°C ±0.2°C over 600°C Tc: R +200 - +1,480°C / ±0.3°C Tc: S +200 - +1,480°C / ±0.3°C Tc: B +200 - +1,800°C / ±0.4°C Tc: E -200 - +750°C / ±0.1°C up to 300°C ±0.2°C over 300°C |
Resolution | 0.05°C in the range 199.95°C 0.1°C in the remaining range |
Temperature drift @20°C | 0.02%/°C |
Drift after 1 year | 0.1°C/year |
Internal watch accuracy | 1min/month max deviation |
Unit of measurement | °C, °F, °K |
Display | LCD backlit graphic display 128x64 pixel. |
Keyboard | 15 keys; the instruments can be
configured without a PC. |
Keyboard locking function | with password. |
Memory | Divided into 64 blocks. |
Storage capacity | Up to 800,000 recordings divided among the existing inputs; for example with one probe connected, 800,000 recordings; with 8 probes connected, 96,000 recordings each. |
Security of stored data | Unlimited. |
Power supply | 4 1.5V alkaline C-BABY type batteries External power supply 12VDC-1A. Connector, external |
Power absorbed @6VDC: | <60μA when the instrument is off <60μA in sleep mode with all probes connected <40mA during data logging with all probes connected |
Autonomy | 200 hours with 7800mAh alkaline
batteries, with all probes connected |
Data download | RS232C from 1200 to 38400 baud,
galvanically isolated. Sub D 9-pole male connector. USB 1.1 - 2.0 galvanically isolated. |
Operating temperature | -5 - 50°C |
Working relative humidity | 0 - 90% RH without condensation |
Protection degree | IP64 |