The DPC-300A Dual System Pressure Calibrator isn’t just a replacement for the old “box” calibrator. With full digital precision and accuracy, a dual pressure system and a built-in loop calibrator function, it’s a pressure calibration powerhouse in a portable format. A dual pressure system means you can use regulated plant air or internally generated pressure sources for calibration. The internal pump goes up to 300 psi with pushbutton ease. Exact values are set using the fine vernier control knob. Plant air, up to 100 psi is controlled with a precision regulator for low inches of water on up.
DPC-300A Dual System Pneumatic Calibrator
Model | Description |
---|---|
DPC-300A | Dual System Pneumatic Pressure Calibrator 300 PSI |
Above DPC-300A Dual System Pneumatic Pressure Calibrator includes:
300 psi/20 bar/200 kPa calibrator, NIST traceable calibration certificate, test leads,
universal AC charger/adapter, hose kit and user guide
300 psi/20 bar/200 kPa calibrator, NIST traceable calibration certificate, test leads,
universal AC charger/adapter, hose kit and user guide
Dual pressure systems
- Regulated plant air to 100 psi / 7 bar.
- Integral electric pump to 300 psi / 20 bar .
Pressure accuracy up to ±0.025% of full scale.
Built-in loop calibrator with 24 V loop power.
External pressure module interface.
24 VDC instrument power supply.
User’s selection of 13 engineering units.
Heavy duty IP67 case (cover closed).
Long life rechargeable NiMH battery power.
- Regulated plant air to 100 psi / 7 bar.
- Integral electric pump to 300 psi / 20 bar .
Pressure accuracy up to ±0.025% of full scale.
Built-in loop calibrator with 24 V loop power.
External pressure module interface.
24 VDC instrument power supply.
User’s selection of 13 engineering units.
Heavy duty IP67 case (cover closed).
Long life rechargeable NiMH battery power.
DPC-300A Pneumatic Calibrator Specification.
Range: Pressure mA Volts |
-12 to 300.00 psi / -0.8 to 20 bar / -80 to 2000 kPa 0 to 24.000 mA 0 to 30.000 VDC |
Engineering Units | psi, bar, mbar, kPa, MPa, kgcm2,
mmH2O @ 4°C, mmH2O @ 20°C, cmH2O @ 4°C, cmH2O @ 20°C, inH2O @ 4°C, inH2O @ 20°C, inH2O @ 60°F, mmHg @ 0°C, inHg @ 0°C |
Media Compatibility: | Clean, non-corrosive gases |
Modes of Operation: | Gauge, Differential, Vacuum (Vacuum to 5 PSI) |
Accuracy: Pressure mA Volts |
6 months: ±0.025% of reading ±0.01% of full scale
1 year: ±0.035% of reading ±0.01% of full scale ±0.015% of reading ±0.002mA ±0.015% of reading ±0.002V |
Operating Range: | -10 °C to +50 °C |
Temperature Coefficient: | 0.01% F.S./°C for temps. outside of 23°C ±5°C |
Storage Temperature: | -20 °C to +60 °C |
Power: | 9 cell NiMH rechargeable; 3700 mAh @ 10.8VApprox.
50 hours (only measurement or (fully-charged) with external pressure supply) 125 pump cycles to 20 bar 300 pump cycles to 10 bar 1,000 pump cycles to 2 bar |
Size/Weight: | 15.25" x 12: x 7" (387.4 x 304.8 x 177.8 mm)/15.5 lbs. (approx. 7 kg) |