5898 High current micro-ohmmeter 200 Amp.
Precision high current portable 200A Micro Ohmmeter.
4 terminal measurement technique. Basic accuracy of 0.2% and an overall
resistance range of 0.1 µΩ to 6 Ω.
High current 200 Amp Micro-ohmmeter
Model | Description |
---|---|
5898 | 5898 High current Micro-ohmmeter. 200 Amp. |
The 5898 Micro Ohmmeter can be used fully auto-ranging or with manual range selection if required.
Each of the 6 ranges has a choice of high or low measuring current.
A four terminal measurement principle to eliminate lead resistance.
The 200A current is pulsed with a duty cycle of approximately 10%. Measurement value displayed
continuously.
Micro ohmmeter specification
High current 200 Amp micro-ohmmeter | |||||
Range | Resolution | Current | Accuracy | Temperature Coefficient (%reading + last digit) |
|
% reading | Last digits | ||||
600 µΩ | 0.1 µΩ | 200A/50A | +/-0.4% | +6 | (0.04% +0.2)/ ° C |
6000 µΩ | 1.0 µΩ | 200A/50A | +/-0.2% | +6 | (0.02% +0.1)/ ° C |
60 mΩ | 10 µΩ | 20A/5A | +/-0.2% | +6 | (0.02% +0.1)/ ° C |
600 mΩ | 0.1 mΩ | 2A/0.5A | +/-0.2% | +6 | (0.02% +0.1)/ ° C |
6 Ω | 1.0 mΩ | 200mA/50mA | +/-0.4% | +6 | (0.04% +0.1)/ ° C |