TC100 Thermocouple Calibrator
The TC100 Thermocouple Calibrator
provides high accuracy source and measurement of ten common
thermocouples, as well as mV.
The TC-100's Thermocouple Calibrators accuracy of ±
0.3°
C for
Type J T/Cs includes all errors,
at resolutions of ±
0.01°
C or °
F in measure mode and ±
0.1°
C or °
F
in source mode.
Features including MIN/MAX recall in measure mode, three set
points per thermocouple range, a large knob
for decade control of the output in source mode, and the ability
to accept bare T/C wires in addition to
mini-plug inputs, makes the TC-100 thermocouple calibrator an
accurate, easy-to-use instrument for all
your thermocouple calibration needs.
TC100 Thermocouple Calibrator
Model | Description |
---|---|
TC100 | Thermocouple
Calibrator includes Protective Sleeve Case, Batteries, Instruction Manual, NIST Calibration Certificate |
Output Voltage | |
Range: | -10 to +75.000 mV |
Resolution: | 1 µV |
Accuracy: | ± 0.007 % of reading, ± 10 µV |
Output Impedance: | ≤1 Ohm |
Input Voltage | |
Range: | -10 to +75.000 mV |
Resolution: | 1 µV |
Accuracy: | ± 0.007 % of reading, ± 10 µV |
Output Impedance: | Input Impedance > 1 MegOhm |
Thermocouple Source/Measure: | |
Types: | J, K, T, E, R, S, B, L, U, C |
Range: | mV |
Resolution | |
Source: | ± 0.1° C or ° F |
Measure | ± 0.01° C or ° F |
Accuracy: | |
J | ±
0.5 °
C; -210 °
C to -100 °
C ± 0.3 ° C; -100 ° C to +1,200 ° C |
K | ±
0.6 °
C; -200 °
C to -100 °
C ± 0.35 ° C; -100 ° C to +1,000 ° C ± 0.5 ° C; +1,000 ° C to +1,372 ° C |
T | ±
0.7 °
C; -200 °
C to -150 °
C ± 0.3 ° C; -150 ° C to +400 ° C |
E | ±
0.5 °
C; -200 °
C to -100 °
C ± 0.3 ° C; -100 ° C to +1,000 ° C |
R | ±
1.8 °
C; 0 °
C to 250 °
C ± 1.0 ° C; 250 ° C to +1,767 ° C |
S | ±
1.8 °
C; 0 °
C to 250 °
C ± 1.0 ° C; 250 ° C to +1,767 ° C |
B | ±
1.7 °
C; 600 °
C to 1,000 °
C ± 1.2 ° C; 1,000 ° C to 1,820 ° C |
L | ±
0.5 °
C; -200 °
C to -100 °
C ± 0.4 ° C; -100 ° C to +900 ° C |
U | ±
0.7 °
C; -200 °
C to 0 °
C ± 0.3 ° C; 0 ° C to +600 ° C |
C | ±
0.4 °
C; 0 to °
C 1,000 °
C ± 0.7 ° C; 1,000 ° C to +1,800 ° C ± 1.2 ° C; +1,800 ° C to +2,316 ° C |
CJC Temperature Offset: | ± 0.05 ° C/° C outside of 23± 5 ° C |
Warm-up Time: | 1 minute to specification |
Environmental: | |
Operating: | 10 ° C to +55 ° C |
Storage: | -20 ° C to +70 ° C |
Power Requirements: | |
Battery: | 9 Volt alkaline; 006P/ IEC 6F22/ NEDA1604 Optional NiCad Optional AC adapter/charger |
Mechanical Dimensions: | 5.7" x 3.15" x 1.43" (14.47 cm x 8.00 cm x 3.63 cm) |
Weight: | 12 ounces (340 grams) |