TC100 Thermocouple Calibrator
The TC100 Thermocouple Calibrator
provides high accuracy source and measurement of ten common
thermocouples, as well as mV.
The TC-100's Thermocouple Calibrators accuracy of ±
0.3°
C for
Type J T/Cs includes all errors,
at resolutions of ±
0.01°
C or °
F in measure mode and ±
0.1°
C or °
F
in source mode.
Features including MIN/MAX recall in measure mode, three set
points per thermocouple range, a large knob
for decade control of the output in source mode, and the ability
to accept bare T/C wires in addition to
mini-plug inputs, makes the TC-100 thermocouple calibrator an
accurate, easy-to-use instrument for all
your thermocouple calibration needs.
TC100 Thermocouple Calibrator
| Model | Description |
|---|---|
| TC100 | Thermocouple
Calibrator includes Protective Sleeve Case, Batteries, Instruction Manual, NIST Calibration Certificate |
| Output Voltage | |
| Range: | -10 to +75.000 mV |
| Resolution: | 1 µV |
| Accuracy: | ± 0.007 % of reading, ± 10 µV |
| Output Impedance: | ≤1 Ohm |
| Input Voltage | |
| Range: | -10 to +75.000 mV |
| Resolution: | 1 µV |
| Accuracy: | ± 0.007 % of reading, ± 10 µV |
| Output Impedance: | Input Impedance > 1 MegOhm |
| Thermocouple Source/Measure: | |
| Types: | J, K, T, E, R, S, B, L, U, C |
| Range: | mV |
| Resolution | |
| Source: | ± 0.1° C or ° F |
| Measure | ± 0.01° C or ° F |
| Accuracy: | |
| J | ±
0.5 °
C; -210 °
C to -100 °
C ± 0.3 ° C; -100 ° C to +1,200 ° C |
| K | ±
0.6 °
C; -200 °
C to -100 °
C ± 0.35 ° C; -100 ° C to +1,000 ° C ± 0.5 ° C; +1,000 ° C to +1,372 ° C |
| T | ±
0.7 °
C; -200 °
C to -150 °
C ± 0.3 ° C; -150 ° C to +400 ° C |
| E | ±
0.5 °
C; -200 °
C to -100 °
C ± 0.3 ° C; -100 ° C to +1,000 ° C |
| R | ±
1.8 °
C; 0 °
C to 250 °
C ± 1.0 ° C; 250 ° C to +1,767 ° C |
| S | ±
1.8 °
C; 0 °
C to 250 °
C ± 1.0 ° C; 250 ° C to +1,767 ° C |
| B | ±
1.7 °
C; 600 °
C to 1,000 °
C ± 1.2 ° C; 1,000 ° C to 1,820 ° C |
| L | ±
0.5 °
C; -200 °
C to -100 °
C ± 0.4 ° C; -100 ° C to +900 ° C |
| U | ±
0.7 °
C; -200 °
C to 0 °
C ± 0.3 ° C; 0 ° C to +600 ° C |
| C | ±
0.4 °
C; 0 to °
C 1,000 °
C ± 0.7 ° C; 1,000 ° C to +1,800 ° C ± 1.2 ° C; +1,800 ° C to +2,316 ° C |
| CJC Temperature Offset: | ± 0.05 ° C/° C outside of 23± 5 ° C |
| Warm-up Time: | 1 minute to specification |
| Environmental: | |
| Operating: | 10 ° C to +55 ° C |
| Storage: | -20 ° C to +70 ° C |
| Power Requirements: | |
| Battery: | 9 Volt alkaline; 006P/ IEC 6F22/ NEDA1604 Optional NiCad Optional AC adapter/charger |
| Mechanical Dimensions: | 5.7" x 3.15" x 1.43" (14.47 cm x 8.00 cm x 3.63 cm) |
| Weight: | 12 ounces (340 grams) |